Manual preset
SELECT PRESET TYPE
Manual
Ext Default
Select Manual with the SET knob and press
SELECT key.
The AK 95 S Presets are based on the concept of
groups. The main groups are:
• 0 Start up
• 1 Configuration
• 2 Set values and limits
• 3 Override times
• 4 Auto priming
• 5 Rinse Back
• 6 Disinfection
• 7 Single needle
• 8 Acetate concentrates
• 9 Bicarbonate concentrates
• 10 BPM setting
• 11 Profiling
Each main group contains a different number of
sub groups.
How to preset manualy
1. Select main group with the SET knob.
2. Press
3. Select sub group with the SET knob.
4. Press
5. Select preset value with the SET knob.
START
6. Press
to store the new value.
UF
STOP
7. Continue with next preset.
Note!
Use
to step back in the preset menue.
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005
External Preset
SELECT PRESET TYPE
Manual
Ext Default
Select Ext with the SET knob and press
SELECT key and following display shows up:
SELECT EXTERNAL PRESET TYPE
PC
Send
External preset in AK 95 S includes both preset-
ting the machine from a PC connected to the
machine (PC preset), and presetting one
machine with the contents of another (duplicate
preset)
This set up is used for customising (within
certain limits) AK 95 S to each clinic's wishes.
For more information, contact Gambro
Technichal Service.