Calibration Of Linear Transducers; Selection Diagram - Gambro AK 95 S Service Manual

Dialysis machines
Table of Contents

Advertisement

Calibrations
Transducers
Calibration of linear
transducers
1.
Start the AK 95 S, enter the Service
menu and select Calibrations.
2.
Choose Transducers
3.
Follow the instructions for each
calibration.
Select Extern for no flow
(pumps stopped).

Selection diagram

VENOUS PRESSURE TRANSDUCER CPU P
VENOUS PRESSURE TRANSDUCER CPU C
ARTERIAL PRESSURE TRANSDUCER
CONDUCTIVITY A TRANSDUCER
CONDUCTIVITY B TRANSDUCER
6:4
PRIMING DETECTOR
BLOOD LEAK DETECTOR
Select Internal and the AK 95 S main-
tains a flow during the calibration.
Independently wich flow support is chosen
the Selection diagram turns up.
4.
When turning the SET knob all the
different transducers that are possible to
calibrate will "scroll" on the Information
display.
Run Time Meter
pH TRANSDUCER
HPG PRESSURE TRANSDUCER
DEGASSING PRESSURE TRANSDUCER
DIALYSIS PRESSURE TRANSDUCER
CONDUCTIVITY A B P TRANSDUCERS
CONDUCTIVITY P TRANSDUCER
AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005