Service Technicians Guide
The flow paths
The flow paths for fluid and blood are shown in different colours on the panel:
Dialyzer
path
Blood path
The blood path ligths up (red) as soon as the
machine detects blood. Locks in blood mode
after 20 seconds.
Bypass path
The bypass path is lit when the dialysis fluid is
manually or automatically bypassing the dia-
lyzer.
During certain alarm situations, and during the
self-calibration of the UF-system, the machine
will automatically bypass the dialyzer.
Red
Green
Red
Orange
4:4
Blood path
Bypass path
Treatment
Green
Green
Treatment, fluid NOT OK
Yellow
Orange
Ultrafiltration
path
Dialyzer path
The dialyzer path is lit when dialysis fluid
passes through the dialyzer.
Colours:
Green = Fluid is correct
Orange = Fluid incorrect
(wrong Temp and/or Cond)
Ultrafiltration path
The ultrafiltration path is yellow during isolated
UF and during bypass.
Treatment, fluid bypass, Isolated UF
Red
Yellow
Green
Standby, fluid OK/NOT OK
Yellow
Green/Orange
Green/Orange
Bypass
Green
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005