Blood pressure control
Venous pressure:
Operating range
Alarm limits:
Accuracy:
Blood Pressure Monitor (BPM)
Systolic pressure range
Low alarm limit, can be
preset (Default)
High alarm limit, can be
preset (Default)
Diastolic pressure range
Low alarm limit, can be
preset (Default)
High alarm limit, can be
preset (Default)
Mean pressure range
Low alarm limit, can be
preset (Default)
High alarm limit, can be
preset (Default)
Pulse rate range
Low alarm limit, can be
preset (Default)
High alarm limit, can be
preset (Default)
1
Meets ANSI/AAMI SP-10 (1992). Mean error ±5 mmHg. Standard
deviation 8 mmHg
H CEN9369 Revision.10.2005
Program version 12.xx
1
1
1
AK 95 S Service m anual - Technical data and spec ifi cations
-700 to 750 mmHg
10 to 500 mmHg in treatment mode, -100
to 500 mmHg in priming mode
-700 to -500 mmHg, ±10% -500 to 500
mmHg, ±5 mmHg or ±3%, whichever is
largest 500 to 750 mmHg, ±10%
60 - 250 mmHg
60 - 250 mmHg (100 mmHg)
60 - 250 mmHg (150 mmHg)
40 - 200 mmHg
40 - 200 mmHg (60 mmHg)
40 - 200 mmHg (100 mmHg)
45 - 235 mmHg
45 - 235 mmHg (75 mmHg)
45 - 235 mmHg (115 mmHg)
40 - 200 bpm (±2 bpm or ±2% of reading)
40 - 200 bpm (40 bpm)
40 - 200 bpm (150 bpm)
3:3