Intern
1. Start the AK 95 S with the main switch and
the ON/OFF button.
2. Place B-pickup tube in the acetate-concentrate
container.
3. Enter Service mode.
4. Select Calibration.
5. Select Transducers.
6. Select UF.
7. Select Intern.
The internal UF calibration runs automatically,
until the calibration is finished.
AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005
When the internal UF calibration is ready:
After the calibration
, A K 9 5 S d i s p l a y s :
INTERNAL UF CALIBRATION READY
SELECT for info, HOLD to save
Press
to show more information.
Press
to save the calibrations values.
When the storage is completed, AK 95 S displays:
EEPROM STORAGE COMPLETED
Press HOLD to exit
When ready, press
calibration.
Note! If the machine is equipped with a UFDkit,
change preset (1-26 UFD) to "not inst" before
entering calibration mode.
Calibrations
to exit the internal UF
6:19