Replacing valve membranes
Correct assembly procedure of the housing and
membrane is essential for proper function and
preventing future leaks.
• Place the membrane in the valve housing with
the seam on the periphery of the silicone
membrane facing from the magnet of the valve.
Valve mounting tool K15391C
A special service tool for mounting of the valve
membrane correctly in the valve housing is
available.
5:20
•Tighten the screws with a torque of
approximatly 0,7 Nm. Make sure that self-
locking nuts are used.
Note! Do not overtighten the screws, this will
deform the valve housing and cause leakage.
B-and C-kit
AK 95 / AK 95 S HCEN12189
Rev.10. 2006