Connection of external equipment
External connector:
External equipment:
Extern alarm:
Max voltage:
Max current:
RS-232C:
Max input voltage:
High level min output
voltage:
Low level max output
voltage:
Max output current:
RS-422/Current loop:
Typ current:
Max current:
H CEN9369 Revision.10.2005
Program version 12.xx
AK 95 S Service m anual - Technical data and spec ifi cations
25 pin D-Sub with RS-232C or
RS-422/Current loop. Opto insulated
fulfilling IEC 60601-1-1.
External equipment intended for
connection to signal input, signal output
or other connectors, shall comply with
relevant IEC standard (e.g. IEC 60950
for IT equipment and the IEC 60601
series for medical electrical equipment).
In addition, all such combinations -
systems - shall comply with the standard
IEC 60601-1-1, Safety requirements for
medical electrical systems. Equipment
not complying to IEC 60601 shall be
kept outside the patient environment, as
defined in the standard.
who connects external equipment to
signal input, signal output or other
connectors has formed a system and is
therefore responsible for the system to
comply with the requirements of IEC
60601-1-1. If in doubt, contact qualified
technician or your local representative.
The normal distance can be mentioned
i.e. at least 1.5 m from the patient or the
patient support.
24 V
AC or DC
100 mA
AC or DC
±15 V
DC
5.0 V
DC
-5.0 V
DC
±5 mA
DC
20 mA
DC
50 mA
DC
1)
Any person
1)
3:9