Service Technicians Guide
Logging mode flow diagram
Display the logging parameters on the
information display see e.g. below:
LOGGING ORI-001
DISPLAY
This logging setup is automatically
stored and shows up next time
internal logging starts.
The logging setups is
automatically stored
in SAFERAM.
4:26
SELECT LOGGER FACILITY
Internal Errors
See next page,
Error buffers in
Logging mode
INTERNAL LOGGING SELECTION
Display Set-up A
ORB-000
ORB-000
xxx
xxx
SET-UP PAR-INH
Array
On
Display
Parameter
name
inhibit.
("ON or
"OFF")
FIELD
ARRAY
0
SRI
Time
Progver
Set-up M
xxx
FIELD
0
Use the Set knob when you want to
Select infor-
set the parameters.
mation display
Setting PAR-INH (PARameter
field (0-2)
INHibit) to ON prevents the logging
display from disappearing if the
Blood Flow knob is touched or a
parameter button (Temp, Cond, etc)
is pressed.
INDEX
FORMAT
000
1
When selecting logging
this information will be
displayed:
Shows program
version.
Shows total
run time
(See "Memory")
Determines position of decimal
point (or displays in HEX). Not
necessary for S-arrays, see
"Arrays", page 4:28
Select array index (000 - 999).
Select the array to be displayed
(see "Arrays", page 4:27)
In which field the logged
parameter will appear.