Base-kit
How to exchange the parts included in
the Base-kit
1.
First perform a combined heat disinfection
program with CleanCart C or liquid citric
acid.
Then perform a heat disinfection program
in combination with CleanCart A or a
Chemical disinfection program with sodium
hypochlorite. Refer to the AK 95 or AK 95 S
Operator's Manual.
2.
Change air filter, K23734001, in the
AC/DC - power supply unit.
3.
Change air filter, K12966001, in the cover
of AK 95 / AK 95 S.
AK 95 / AK 95 S HCEN12189 Rev.
10. 2006
05
06
07
09
10
08
4.
Change concentrate filters, K16538A, on
the A and B-concentrate connectors.
5.
Change o-rings, 100319061, on the A and
B-concentrate connectors (male).
100319061
6.
Change o-rings, 100319060, O-ring, A- and
B-connectors (female)
100319060
Maintenance Manual