Gambro AK 95 S Service Manual page 246

Dialysis machines
Table of Contents

Advertisement

ELT - Earth Leakage Current Test
File the highest earth leakage current value when the machine is running
according to section 4.3:
Earth leakage current value: _______ µA
Note!
Check that the highest measured readings don't exceed the limit values
described in section 4:3
PLT - Patient Earth Leakage Current Test
File both the highest AC-reading and the highest DC-reading when the
machine is running according to section 4.4:
AC-reading: _______ µA
DC-reading: _______ µA
Note!
Check that the highest measured readings don't exceed the limit values
described in section 4:4
Date:__________________________
Signature:_______________________
Other comments:
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
11:12
Electrical Safety Inspections for AK-machines
HCEN12190 Rev 05.2006