Gambro AK 95 S Service Manual page 68

Dialysis machines
Table of Contents

Advertisement

Bicarbonate:
Measuring range:
Alarm limits:
pH-meter:
Accuracy:
Alarm limits:
Ultrafiltration control
Volume control:
UF volume:
Accuracy of measured
volume:
UF coefficient :
UF rate:
Time:
TMP:
Alarm limits:
Accuracy:
Blood leakage detection
Detection method:
Sensitivity:
Disinfection and cleaning
Chemical disinfection:
Peracetic Acid Program (presetable)
Concentration of disinfectant
Concentration in machine
Volume
Contact time between treatments
Contact time overnight or when not in
use (recommended with UFD-kit)
Total time
AK 95 HCEN9135 Rev. 10.2005
+
Na
, 130 to 160 mmol/l, (±6 mmol/l)
HCO
- , 20 to 40 mmol/l, (±6 mmol/l)
3
12 to 16 mS/cm
To 5 % of the calculated conductivity set value
1 – 9.9 pH, measured in the dialysis fluid before dialyzer.
±0.2 pH
5.0 to 9.0 pH
Direct electromagnetic measurement of dialysis fluid flow,
before and after the dialyzer.
Adjustable 0 to 10.00 l
±50 ml/h or ±1 % whichever is largest
Maximum 85 ml/h/mmHg
0.0 to 4 l/h, given by the set values of UF volume and
treatment time.
Remaining treatment time control, adjustable up to 9 h 59
min.
TMP is defined as the difference, P
is the venous pressure and P
the dialysis fluid, where it enters the machine after the
dialyzer.
-100 to 500 mmHg
±500 ( ±10 mmHg or ±5 %, whichever is largest )
Infrared light detector.
Alarm will be given for ≥0.3 ml blood, haematocrit 32%, per
minute at 700 ml/min dialysis fluid flow.
Time delay for alarm, maximum 5 seconds (diffusion mode).
3.5 % peracetic acid
0.1 %; i.e. diluted 1 + 34 (1:35)
approx. 85 ml (with UFD-kit approx. 92 ml)
10 minutes
Minimum 3h dwell time
30 minutes (with UFD-kit 59 minutes)
including 10 min contact time
- P
, where P
b out
d out
is the pressure measured in
d out
Technical Data
b out
3:5