Gambro AK 95 S Service Manual page 65

Dialysis machines
Table of Contents

Advertisement

Technical Data
Machine specification AK 95, K18022
Blood flow control
Double needle
Flow rate:
Flow accuracy:
Accumulated blood volume:
Volume accuracy:
Single needle
Flow rate:
Flow accuracy:
Time control:
Pressure control:
Accumulated blood volume:
Volume accuracy:
Blood pressure supervision
Venous pressure:
Operating range
Alarm limits:
Accuracy:
Arterial pressure:
Operating range
Alarm limits:
Accuracy:
3:2
0 and 20 - 500 ml/min (-150 mmHg prepump pressure,
pump segment diameter 7.9 mm)
±10 ml/min or ±10 %, whichever is largest
0 - 999 litres
±10 %
0 and 20 - 500 ml/min (-150 mmHg prepump pressure,
pump segment diameter 7.9 mm)
±10 ml/min or ±10 %, whichever is largest
2 - 20 seconds, arterial and venous time.
10 - 500 mmHg ( ±5 mmHg or ±3 %, whichever is largest),
venous pressure control
0 to 999 litres
±10 %
-700 to 750 mmHg
10 to 500 mmHg in treatment mode,
-100 to 500 mmHg in priming mode
-700 to -500 mmHg, ±10 %
-500 to 500 mmHg, ±5 mmHg or ±3 %, whichever is largest
500 to 750 mmHg, ±10 %
-700 to 750 mmHg
-700 to 750 mmHg
-700 to -500 mmHg, ±10 %
-500 to 500 mmHg, ±5 mmHg or ±3 %, whichever is largest
500 to 750 mmHg, ±10 %
AK 95 HCEN9135 Rev. 10.2005