Service Technicians Guide
UF diagnostics
DIAGNOSTIC FM MENU
Var flow
Pumps UF
Select "UF" with the SET knob and press the
SELECT key and following display appears:
SELECT UF CELL TARATION
Complete
Select "Complete" with the SET knob and press
the SELECT key and following display appears:
COMPLETE UF CELL TARATION
Please wait
A complete taration as in treatment is now
performed. When ready following display occurs:
COMPLETE UF CELL TARATION READY
4:24
Valves diagnostics
DIAGNOSTIC FM MENU
Var flow
Valves
Select "Valves" with the SET knob and press the
SELECT key and following display appears:
INLET VALVE (INVA)
Press SELECT key to toggle valve
Turn the SET knob and all the different valves
in the machine that are possible to turn on and
off will "scroll" on the Information display.
When the SELECT key is pressed the selected
valve toggles on or off. When the valves not are
selected they are off.
INLET VALVE
AIR INLET VALVE
RECIRCULATION VALVE
FLUSHING VALVE
RINSING VALVE
ZEROING VALVE
DIRECT VALVE CPU C
DIRECT VALVE CPU P
TARATION VALVE
EVACUATION VALVE
BYPASS VALVE CPU C
BYPASS VALVE CPU P
Note! Be careful when open the AIVA to avoid
Fluid leakage into the monitor.
Pumps UF
Valves
INVA
AIVA
REVA
FLVA
RIVA
ZEVA
DIVA
DIVA
TAVA
EVVA
BYVA
BYVA
AK 95 / AK 95 S Service Manual