Service Technicians Guide
5. Place the alphanumerical overlay,
K1 5824 001, on the operatorspanel.
6. Edit correct values, press
parameter.
START
7. Press
when all parameters are done
UF
STOP
to load the values into the EEPROM.
8. Press
to exit.
Following parameters are available in the concentrat preset:
Acetate concentrate
Ac
Bicarbonate concentrate
A
B
BiCart
4:14
after each
gram/l - molecule
Label
CH
COONa
3
H
O with CH
COONa
2
3
NaCl
KCl
MgCl
2
H
O with MgCl
2
2
CaCl
2
H
O with CaCl
2
2
Label
Acid (CH
COOH)
3
CH
COONa
3
H
O with CH
COONa
2
3
NaCl
KCl
MgCl
2
H
O with MgCl
2
2
CaCl
2
H
O with CaCl
2
2
NaHCO
3
NaCl
CH
COONa
3
H
O with CH
COONa
2
3
BiCart
A a
K k
B b
C c
O o
D d
E e
S s
F f
G g
W w
I i
H h
Y y
J j
SHIFT
PRESETOVERLAY AK90s
mmol/l - molecule
Label
CH
COONa
3
NaCl
KCl
MgCl
2
CaCl
2
Label
Acid (CH
OOH)
3
CH
COONa
3
NaCl
KCl
MgCl
2
CaCl
2
NaHCO
3
NaCl
CH
COONa
3
BiCart
L l
M m
N n
P p
Q q
R r
T t
U u
V v
X x
Z z
mmol/l - ions
Label
+
Na
K
+
Mg
2+
2+
Ca
-
Cl
CH
COO
-
3
Label
+
Na
K
+
Mg
2+
2+
Ca
Cl
-
CH
COO
-
3
+
H
+
Na
HCO
-
3
Cl
-
-
CH
COO
3
BiCart
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005
0 = 1 &
2 "
3 /
4 # 5 +
6 %
7 -
8 _
9
.

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Gambro AK 95 S

This manual is also suitable for:

Ak 95

Table of Contents