System and Treatment
Parameters
The system and treatment parameters can be
set at five different levels (table 1).
Level Type of parameters
4
Treatment parameters
3
Alarm limits and
defaults
2
Clinic or customer
adaptions
1
Market adaptions
0
Safety limits
Table 1. System and Treatment parameters
For information about the use and functions of
the parameters in level 2 and 3, see the chapter
Technicians Guide.
For information about the use and functions of
the parameters in level 4, see
AK 95 Operators Manual, HCEN 9542 or
AK 95 S Operators Manual, HCEN 9794.
AK 95 / AK 95 S Service Manual
HCEN 9134 Rev. 10.2005
Adjustable by...
Operator, nurse
Clinics technician
Gambro technician or
Gambro subsidiary
Gambro Head office
Fixed
Technical Description