Gambro AK 95 S Service Manual page 240

Dialysis machines
Table of Contents

Advertisement

4
4.1
Safety tester:
According to IEC 60601-1
Suggested Manufacturer: Metron, Rigel, Biotec
4.2
1.
2.
3.
4.
5.
11:6
ELT / PLT
Test equipment
General conditions for ELT / PLT
The AK-monitor shall be connected to both feed water and drain.
Connect the AK-monitor to the outlet supply of the safety tester.
Use a mains plug adapter for respective mains plug. If the AK-
machine is equipped with two mains plugs, the measurements
shall be performed for each mains plug and the values shall be
added.
Feed water
AK-machine
Drain
Central concentrate delivery systems should not be used. Use
either liquid or non-liquid concentrates.
No other external equipment than specified in this instruction
should be connected to the machine.
The protective earth of the machine must not be in contact with
any external protective earth.
Electrical Safety Inspections for AK-machines
Safety
tester
mains
prot. earth
leakage current
earth enclosure
leakage current
alternative
direct
differential alternative
equipment
patient
patient leakage current
lead to GND lead to lead
AC/DC
lead ISO
outlet power
Inlet Power
open E. open N.
Mains voltage to
the AK-machine
230 VAC
ev. M.
HCEN12190 Rev 05.2006