Gambro AK 95 S Service Manual page 52

Dialysis machines
Table of Contents

Advertisement

Ultrafiltration control
Volume control:
UF volume:
Accuracy of measured
volume:
UF coefficient:
UF rate:
Time:
Profiling
UF-rate
+
Na
+
Na
HCO
Disinfection and cleaning
Chemical disinfection
Peracetic Acid Program (presetable)
Concentration of disinfec-
tant
Concentration in machine
Volume
Contact time between
treatments
Contact time overnight or
when not in use (recom-
mended with UFD-kit)
Total time
Hypochlorite Program (presetable)
Concentration of disinfec-
tant
Concentration in machine
Volume
Contact time between
treatments
Contact time overnight or
when not in use
Total time
H CEN9369 Revision.10.2005
Program version 12.xx
, Acetate mode
, Bicarbonate mode
-
, Bicarbonate mode
3
AK 95 S Service m anual - Technical data and spec ifi cations
Direct electromagnetic measurement of
dialysis fluid flow, before and after the
dialyzer.
Adjustable 0 to 10.00 l
±50 ml/h or ±1% whichever is largest
Maximum 85 ml/h/mmHg
0.0 to 4 l/h, given by the set values of UF
volume and treatment time.
Remaining treatment time control. 0.05
to 9.59 hour. minute (±1 minute)
0.0 to 4 l/h
130 to 160 mmol/l
130 to 160 mmol/l
20 to 40 mmol/l
3.5% peracetic acid
0.1%, i.e. diluted 1 + 34 (1:35)
approx. 85 ml (with UFD-kit approx. 92
ml)
10 minutes
Minimum 3h dwell time
30 minutes (with UFD-kit 59 minutes)
including 10 min. contact time
10% available chlorine
0.5%, i.e. diluted 1 + 19 (1:20)
Approx. 145 ml (with UFD-kit approx.
153 ml)
10 minutes
Maximum 20 min, not intended for
overnight disinfection!
30 minutes (with UFD-kit 48 minutes)
including 10 min. contact time
3:5