Blood leak detector

The blood leak detector includes the following
components:
Housing with cover.
Transmitter and Receiverboard
Blood leak detector board.
The transmitter and receiver boards are con-
nected to the blood leak detector board, via
twisted cables. The blood leak detector board
interfaces directly to the FM I/O board, via a
board connector.

Suction pump

The suction pump consists of a gearwheel pump
driven by a DC-motor.
MM I/O board
The MM I/O board is mounted in the mixing
module. It transfers control and data informa-
tion received from the CPU boards to the various
electronic and electromechanical parts in the
mixing module. It also transmits sensor informa-
tion from the mixing module up to the CPU
boards in the AK 95 S.
AK 95 / AK 95 S Service Manual
HCEN 9134 Rev. 10.2005

CPU board

AK 95 S is equipped with one Control CPU board
and one Protective CPU board.
The CPU boards comprises among other things
the microcomputers. The spare parts list gives
the board connectors which reside on the CPU
boards.

Battery

The battery is a rechargeable 9V battery. The
battery is used to drive the buzzer alarm during
mains power failure. The battery is not con-
nected at delivery.
NOTE! The battery must be connected during
the installation according to the instruction in
the Installations Guide.

Fan

The drive motor (DC-motor) and the fan form a
single unit. The fan is connected to the driver
board, via a cable.

Driver board

AK 95 S is equiped with one Control Driver
board and one Protective Driver board, according
to the figure 8.