Connection To The Water Supply - Gambro AK 95 S Service Manual

Dialysis machines
Table of Contents

Advertisement

Installation Guide
Connection to the Water
Supply
The water must fulfil all requirements set out
under Environment and Physical Data.
Connection to the Power
Supply
Mains power,
connected to a
grounded
power socket
1:4
Water in
Drain
Fuses
Main switch,
ON/OFF
Connection of the Battery
The rechargeable 9V battery (100 224 006) is
used to drive the buzzer alarm during main's
power failure. The battery is not connected at
delivery.
NOTE! The battery must be connected during
the installation according to the instruction on
next page.
A discharged battery will be charged with
maximum 10 mA and it must be charged for at
least 50 hours to get completely charged. An all-
up charged battery will receive a maintaining
current of approximately 1,5 mA. The AK 95 S
charge the battery as long as the AC/DC con-
verter is switched on, i.e. AK 95 S itself does not
need to be switched on.
The battery is tested during every functional
check to ensure that the capacity is enough in
the battery, otherwise an attention will be
given:
BATTERY FAILURE
Change battery or press SELECT
NOTE! Jumper J200 on the Protective Driver
board must be mounted for the rechargeable
battery.
If use of an alkaline non-chargeable battery,
jumper J200 must be removed, if not, the
battery will be damaged.
AK 95 / AK 95 S Service Manual
HCEN 9133 Rev. 01. 2004