Gambro AK 95 S Service Manual page 71

Dialysis machines
Table of Contents

Advertisement

Technical Data
Mains plug:
Maximum earth leakage
current:
Maximum patient leakage
current:
Note:
Connection of external equipment
External connector:
External equipment:
Extern alarm:
RS-232C:
RS-422/Current loop :
3:8
For 100/110V, Hospital grade, earth plug, type IEC 83 A5-
15
For 230V, Dual earth system plug, type IEC 83 C4
100 µA (100V)
150 µA (110V)
205 µA (230V, 50 Hz )
400 µA (230V, 60 Hz )
9 µA (100V)
11 µA (110V)
20 µA (230V)
All leakage currents specified are without external equipment
connected to the AK 95.
25 pin D-Sub with RS-232C or RS-422/Current loop.
Opto insulated fulfilling IEC 60601-1-1.
External equipment intended for connection to signal input,
signal output or other connectors, shall comply with relevant
IEC standard (e.g. IEC 60950 for IT equipment and the IEC
60601 series for medical electrical equipment). In addition, all
such combinations – systems – shall comply with the
standard IEC 60601-1-1, Safety requirements for medical
electrical systems. Equipment not complying to IEC 60601
shall be kept outside the patient environment, as defined in
1)
the standard.
Any person who connects external equipment to signal input,
signal output or other connectors has formed a system and is
therefore responsible for the system to comply with the
requirements of IEC 60601-1-1. If in doubt, contact qualified
technician or your local representative.
1)
The normal distance can be mentioned i.e. at least 1.5 m
from the patient or the patient support.
Max voltage:
Max current :
Max input voltage:
High level min output voltage:
Low level max output voltage:
Max output current :
Typ current :
Max current :
24 V
AC or DC
100 mA
AC or DC
±15 V
DC
5.0 V
DC
-5.0 V
DC
±5 mA
DC
20 mA
DC
50 mA
DC
AK 95 HCEN9135 Rev. 10.2005