Gambro AK 95 S Service Manual page 55

Dialysis machines
Table of Contents

Advertisement

Drain:
Length of drain tube:
High pressure guard
transducer:
Inlet pressure switch
(INPS):
Safety guard switch
(SAGS):
Power supply
Mains voltage:
Note: The user must verify the quality of the protective earth in the
installation.
Power consumption:
Cable length:
Cable area:
External fuses:
Mains plug:
Maximum earth leakage
current:
Maximum patient leakage
current:
Note: All leakage currents specified are without external equipment
connected to the AK 95 S.
3:8
AK 95 S Service m anual - Technical data and spec ifi cations
The drain tube outlet must be placed
between floor level and maximum 1.2
m above the outlet connection from the
fluid monitor. An air gap to atmospheric
pressure must always be arranged at the
tube outlet.
<10 m
±500 mmHg, (±10 mmHg or ±5%,
whichever is largest)
Increasing: +150 mmHg, ±20 mmHg
Decreasing: +99 mmHg, ±20 mmHg
Increasing: -59 mmHg, ±7 mmHg
Decreasing: -74 mmHg, ±7 mmHg
100 V (at 50 Hz or 60 Hz) , 110 V (at 50
Hz or 60 Hz) or 230 V (at 50 Hz or 60
Hz) (∼AC), (±12%). The voltage has to
be specified before installation.
2025 W at 230 V 1575 W at 110 V 1275
W at 100 V
3 m
AWG 14 at 100 V and 110 V 1 mm
230 V
For 100/110 V, 2 x 12 AT for heater. For
230 V, 2 x 12 AT for heater.
For 100/110 V, Hospital grade, earth
plug, type IEC 83 A5-15 For 230 V, Dual
earth system plug, type IEC 83 C4
100 µA (100 V) 125 µA (110 V) 200 µA
(230 V)
10 µA (100 V) 10 µA (110 V) 10 µA
(230 V)
2
at
HCEN9369 Revision.10.2005
Program version 12.xx