Gambro AK 95 S Service Manual page 80

Dialysis machines
Table of Contents

Advertisement

Service maintenance menu
Below follows a description of how to use the
buttons in the Control group for preset, logging
and service mode. By way of introduction the
information display together with the two
control buttons are illustrated step by step.
Subsequent information are given in the differ-
ent flow diagrams.
General handling
To get the leading information on the display it
is necessary to press the
hidden buttons above the dialyser symbol at the
same time for 5 sec.
NOTE! The info. display must be dark.
The following will be displayed:
MAINTENANCE SUPPORT SELECTION
Logging
Service/Preset
Select mode with the Set knob and press
confirm.
To enter the Serv/Pres mode a code must be set.
ENTRY PRESET/SERVICE CODE
Code:
XX
Set the code with the Set knob (two digits for
each field). If you do not confirm the code
within 40 seconds, restart from the first figure is
required.
Choose the desired mode by turning the
Set knob.
When desired mode is blinking, confirm
with the
To store values or modes, press
To step backwards and finally exit from
the menu, press
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005
button and two
XX
button.
START
UF
STOP
.
To step through the Service mode, use the Control
buttons as described below.
SET knob and SELECT key, to select.
HOLD key, to step backward.
to
XX
.
Service Technicians Guide
4:7

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Gambro AK 95 S

This manual is also suitable for:

Ak 95

Table of Contents