4.3
1.
2.
3.
4.
5.
6.
Limit values for ELT
AK 95 S (100V)
AK 95 S (110V)
AK 200 S (115V)*
AK 200 ULTRA S (115V)*
AK 95 S (230V)
AK 200 S (230V)
AK 200 ULTRA S (230V)
*The total earth leakage current is the sum of the readings from both
mains cables (to the heater rods and to the ACDC).
HCEN12190 Rev 05.2006
ELT - Earth Leakage Current Test
Let the machine pass FCH.
Measure the earth leakage current with the safety tester.
Take the highest reading when the machine is running.
Invert the phases of the mains voltage. The machine might
restart.
Take the highest reading when the machine is running.
Check that the highest measured readings don't exceed the limit
values in following table:
Max 500 µA
Typical < 100 µA
Max 500 µA
Typical < 125 µA
Max 500 µA
Typical < 100 µA
Max 500 µA
Typical < 100 µA
Max 500 µA
Typical < 200 µA
Max 500 µA
Typical < 145 µA
Max 500 µA
Typical < 145 µA
Electrical Safety Inspections for AK-machines
11:7