Gambro AK 95 S Service Manual page 101

Dialysis machines
Table of Contents

Advertisement

Service Technicians Guide
Arrays
Use the set knob to select the array type.
The arrays used for service purpose are:
SRB
SRD
SRI
SRL
SRR
SRW
The different letters stands for:
First letter Second letter
S=Service
R=RAM
Use the set knob to select arrays and press
to step backwards.
When it is necessary to look at two or three
different arrays at the same time, use the
possibility to choose a separate field on the
information display for each data.
Summary: 1. Set field
2. Set array
3. Set index
4. Push
5. Push "display" to see the logged
parameters.
4:28
Third letter
B=Byte
D=Double Word
I=Integer
L=Logical
R=Real
W=Word
twice to step back
AK 95 / AK 95 S Service Manual