Gambro AK 95 S Service Manual page 58

Dialysis machines
Table of Contents

Advertisement

Performance and specification - Supervisory
system
Blood pressure supervision
Venous pressure:
Operating range
Alarm limits:
Accuracy:
Arterial pressure:
Operating range
Alarm limits:
Accuracy:
Extracorporeal blood loss to the environment
Detection method:
Air detection
Detection method:
Drip chamber size:
Sensitivity:
Extracorporeal blood loss due to coagulation
Detection method:
Alarm:
H CEN9369 Revision.10.2005
Program version 12.xx
AK 95 S Service m anual - Technical data and spec ifi cations
-700 to 750 mmHg
10 to 500 mmHg in treatment mode, -100
to 500 mmHg in priming mode
-700 to -500 mmHg, ±10% -500 to 500
mmHg, ±5 mmHg or ±3%, whichever is
largest 500 to 750 mmHg, ±10%
-700 to 750 mmHg
-700 to 750 mmHg
-700 to -500 mmHg, ±10% -500 to 500
mmHg, ±5 mmHg or ±3%, whichever is
largest 500 to 750 mmHg, ±10%
Venous pressure supervision
Ultrasonic detector placed at the venous
drip chamber. The detector has a two
channel structure and the function of the
detector is tested at the function test made
by the microcomputers.
Diameter 22 mm
Bubbles larger than 1 µl will be trapped
by the drip chamber. An alarm will be
issued if the blood level falls below the
middle of the air detector.
Supervision of the stop time of the blood
pump
The attention "BLOOD PUMP STOP
TIME EXPIRED, Start Blood pump".
3:11