Gambro AK 95 S Service Manual page 85

Dialysis machines
Table of Contents

Advertisement

Service Technicians Guide
Preset parameter description
The presetable parameters for autopriming are
explained in following diagram. The priming
volume as a function of the time for the seven
different phases in autopriming:
• Fill
• Pressure test
• Priming
• Flush 1
• Flush 2
• Level adjustment in the venous drop chamber
• Venous line fill
Priming
volume
Preset
parameter
4:12
Pressure
Venous pressure
alarm limit
Pressure
build time
Pressure test
Priming
Fill
1,2
3,4,5
6,7,8
For instructions how to operate the autopriming
mode, refer to AK 95 S Operatorsmanual.
Before starting the autopriming, it is essential
that correct settings of the preset parameters
have been made, corresponding to the type of
dialyzer and blood lines used.
Make sure that the total volume of priming
solution is enough to cover the preset flow rate
and priming time.
The preset values must be set in accordance
with the respective users needs.
Pressure drop
Pressure
Time
drop time
Level
adjust-
ment
Flush 2
Flush 1
6,10
6,11
6,9
6,12
Venous
line fill
Time
AK 95 / AK 95 S Service Manual
HCEN 9136 Rev. 10.2005