Dialysis fluid preparation
Temperature:
Temperature alarm (fixed)
Overheat protection:
Conductivity:
Alarm limits:
Ultrafiltration supervision
TMP:
Alarm limits:
Accuracy:
Blood leakage detection
Detection method:
Sensitivity:
pH-meter
Measuring range:
Accuracy:
Alarm limits:
3:12
AK 95 S Service m anual - Technical data and specificat
40 °C (±0.5 °C)
The heater can only be on if there is a
flow through the heater. The flow is
detected by a flow switch.
±5 % of the calculated conductivity set
value
TMP is defined as the difference, P
P
, where P
is the venous pressure
d out
b out
and P
is the pressure measured in the
d out
dialysis fluid, where it enters the machine
after the dialyzer.
-100 to 500 mmHg
±500 ( ±10 mmHg or ±5 %, whichever is
largest )
Infrared light detector.
Alarm will be given for ≥0.3 ml blood,
haematocrit 32%, per minute at 700
ml/min dialysis fluid flow. Time delay
for alarm, maximum 5 seconds (diffusion
mode).
1 - 9.9 pH, measured in the dialysis fluid
before dialyzer.
±0.2 pH
5.0 to 9.0 pH
-
b out
HCEN9369 Revision.10.2005
ions
Program version 12.xx