5.2 Preparing for the function check

1. Check battery status: The rechargeable battery must be fully
charged.
If necessary: Charge rechargeable battery.
2. Check the following parts for external damage:
Device
Labels on the device
Connectors and cables
Breathing circuit
Accessories
3. If necessary: Replace damaged parts.
4. Check the patient valve of the reusable breathing circuit
(see "5.4 Checking the reusable breathing circuit", page
5. If necessary: Replace the breathing circuit.
6. Check the fill level of the oxygen cylinder.
7. If necessary: Replace the oxygen cylinder.
Result
The function check has been prepared.

5.3 Performing the function check

You can perform the function check with the following test lungs:
Testing bag WM 1453
Testing bag WM 1454
EasyLung for WEINMANN Emergency WM 28625
MEDUVENT Standard
5 Function check
52).
EN