9

Description of the modes

9.1 CPAP

Pressure
Druck
CPAP / PEEP
CPAP / PEEP
Description
Abbreviation
Long form
Type
Requirement
Ventilation parameters
O
i
2
pMax
CPAP
1/freq. (spontaneous)
1/Freq. (spontan)
The CPAP / PEEP setting increases the lower pressure level during
spontaneous breathing. This results in an increase in the functional
residual capacity of spontaneously-breathing patients. The patient
is able to breathe spontaneously without any restriction at the set
pressure level. The CPAP mode is used exclusively on patients with
adequate spontaneous breathing.
The pressure is generally set at the end of expiration (PEEP).
9 Description of the modes
CPAP
Continuous Positive Airway Pressure
Pressure-controlled
None
MVi
PEEP
CPAP
1/freq. (spontaneous)
1/Freq. (spontan)
MEDUVENT Standard
Time
Zeit
EN