10 Description of the modes

10.2 CPAP

Pressure
Druck
CPAP / PEEP
CPAP / PEEP
122
EN
MEDUVENT Standard
Hazardous therapy as a result of an unsuitable ventilation
mode!
In ventilation modes CPAP and CPAP + PS, patients do not receive
any mandatory mechanical breaths. In patients with inadequate
spontaneous breathing or respiratory arrest, use of these modes
might lead to hypoventilation, and cause the patient serious or
life-threatening injury.
 Use CPAP or CPAP + PS modes only on patients with adequate
spontaneous breathing.
 Continuously monitor the patient and device during ventilation.
Description
Abbreviation
Long form
Type
Requirements
Ventilation parameters
PEEP
CPAP
1/freq. (spontaneous)
1/Freq. (spontan)
The CPAP/PEEP setting increases the lower pressure level during
spontaneous breathing. This results in an increase in the functional
residual capacity of spontaneously breathing patients. The patient
is able to breathe spontaneously without any restriction at the set
pressure level. Pressure is generally set at the end of expiration
(PEEP).
CPAP
Continuous Positive Airway Pressure
Pressure-controlled
None
CPAP
1/freq. (spontaneous)
1/Freq. (spontan)
Time
Zeit