5.4 Selecting emergency mode

Emergency
ventilation
function
1
2
3
Emergency ventilation is started with preset parameters. These parameters can be changed
in the operator menu (see "8.2 Ventilation Modes" on page 99) and are optimized for the
following patient groups:
Infant (up to approx. 1 year, 10 kg body weight)
Child (approx. 1-12 years, 25 - 30 kg body weight)
Adult (approx. 13 years and over, 75 kg body weight)
EN
Operation
46
Three modes with preset ventilation parameters are
available for emergency ventilation. You can select these
directly at any time during ventilation by pressing one of
the function buttons twice or by pressing one of the
function buttons once and then confirming with the
navigation knob.
• Button 1: Infant
• Button 2: Child
• Button 3: Adult
If IPPV is selected as the emergency mode, the display
shows a pressure gauge. If BiLevel + ASB is selected as the
emergency mode, the display shows the most recently
used curve display.
Exit the emergency mode as follows:
• Select a mode in the "Mode" menu
• Select the menu item "Curves" in the main menu.
IPPV and BiLevel + ASB can be selected as emergency
modes in the operator menu (see "Emergency Mode" on
page 107). IPPV mode is always preset at the factory. This
mode is activated automatically when you call up an
emergency mode from another ventilation mode.