Physician menu
3.2 Parameters in the Physician menu
Parameter
Therapy mode
If CPAP: CPAP
pressure
If APAP: lower
pressure limit
If APAP: upper
pressure limit
Rate of pressure
rise (APAP)
softPAP level
softPAP lock
8
EN
Display
on
device
APAP
CPAP, APAP (SOMNObalance (e) only)
CPAP
SOMNObalance / SOMNOsoft 2:
4 hPa - 18 hPa
P
SOMNObalance e / SOMNOsoft 2e :
4 hPa - 20 hPa
P
4 hPa-P
min
SOMNObalance / SOMNOsoft 2:
P
P
max
SOMNObalance e / SOMNOsoft 2e :
P
0. hPa - 0.4 hPa/s
0 (none/off)
Soft
1 (slight drop)
2 (normal drop)
0 (lock disabled, patient can change
SLOC
1 (lock enabled, patient cannot
SOMNObalance (e) / SOMNOsoft 2 (e)
Value range
max
-18 hPa
min
-20 hPa
min
softPAP setting)
change softPAP setting)
Need help?