6.5 Changing ventilation parameters

Requirement
1. Select the ventilation parameter you want to change using the
2. Change the ventilation parameter.
3. Confirm the value.
4. Repeat the steps for all the ventilation parameters you want to
Result
Ventilation parameters have been changed.
The following parameters are mutually dependent:
The device is switched on
page
54).
Ventilation has been started
page
56).
navigation knob.
change.
The minimum difference between pMax and PEEP is 5 mbar.
Frequency /Vti: The product of freq. x Vti results in inspiratory
minute volume MVi.
The ventilation parameters "frequency" and "tidal volume"
can only be set in meaningful combinations. Combinations
resulting in a value below 1.5 l/min or above 20 l/min (BTPS)
cannot be set.