Ventilation Parameters - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

Ventilation
Inspiration trigger
mode
IPPV
No
BiLevel + ASB
Yes
CPAP
No
CPAP + ASB
Yes
aPCV
Yes
PCV
No
PRVC + ASB
Yes
SIMV
Yes, permanently set
SIMV + ASB
Yes
Yes (with Flow
S-IPPV
measurement + ASB
option can be set)
Demand
Yes, permanently set
CCSV
Yes

7.2 Ventilation parameters

There are the following trigger options in the individual ventilation
modes:
Expiration
trigger
No
Yes
No
Yes
No
No
Yes
No
Yes
No
Yes, permanently set No
No
Ventilation
parameters
Vt
Freq.
pMax
PEEP
Air Mix
93 % oxygen
100 % oxygen
ΔpASB (only with
Flow
measurement + ASB
option)
7 Description of the modes
Trigger window
for mandatory
breaths
No
20 % of Te
No
No
Can be set from
0 % - 100 % of Te
No
20 % of Te
20 % of Te
20 % of Te
100 % of Te
No
Unit
Description
ml
Tidal volume (breath volume)
1/min
Ventilation rate
mbar
Maximum ventilation pressure
mbar
Positive end-expiratory pressure (CPAP)
Ventilation through the addition of
-
ambient air
-
Ventilation with concentrator oxygen
-
Ventilation with 100 % oxygen
mbar
Pressure support (relative to the set PEEP)
MEDUMAT Standard
ASB breath
No
Yes
No
Yes
No
No
Yes
No
Yes
No
No
No
2
EN
141

Advertisement

Table of Contents
loading

Table of Contents