Change Of Patient - Weinmann homecare SOMNOsoft + WM 24600 Description And Operating Instructions

Table of Contents

Advertisement

Available languages

Available languages

Hose system
The corrugated hose WM 24120 (grey) can be autoclaved in devices
complying with EN 285. Temperature: 134 °C, minimum residence time
3 minutes.
The corrugated hose WM 24130 (transparent) can be cleaned in water
at up to 70°C. Sterilization is not permitted.
As a disinfectant we recommend GIGASEPT FF.
The procedure for disinfection is the same as for cleaning.
After disinfection, all parts must be thoroughly rinsed with distilled water
and dried.
Allow the hose system to drip dry. Dry the hose system with the
SOMNOsoft +, as described on page 19ff.
Accessories
When disinfecting or sterilizing the exhalation system, mask, bacteria filter and humidifier,
please observe the chapters on "Disinfection and sterilization" in the relevant operating in-
structions.

6.4 Change of patient

If the device is to be used for another patient without using a bacteria filter, it must first
undergo a hygiene procedure. This must carried out by Weinmann or a specialist dealer.
The details of the hygiene procedure are described in the
instructions.
22
Hygienic preparation
SOMNOsoft
+ service and repair