Exchanging the Diverter Valve

Frequency
Tools required
Parts required
Before beginning this procedure:
Position the transport unit of the fraction
collector in the "Home Position" (see
"Maintenance Functions"
Remove all installed trays from the tray
base.
Position the transport unit of the fraction
collector in the "Change Parts Position"
(see
"Maintenance Functions"
Turn off the instrument.
2 Unscrew the 2 finger tight fittings of the inlet
/ waste tubing assembly at the diverter valve.
Agilent 1100 Series Fraction Collectors User's Guide
When leaky or defective
Hex key 2.0 mm, PN 8710-2438 (supplied in accessory kit)
Diverter Valve, PN G1364-61901
on page 49).
on page 49).
Repairing the Fraction Collector
1 Locate the diverter valve with the finger tight
fittings of the inlet / waste tubing assembly
(the figure shows the open fraction collector
seen from the front).
3 Unscrew the finger tight fitting of the valve to
needle tubing assembly at the diverter valve.
3
Simple Repairs
71