Initialization Failed - Agilent Technologies G1313A Reference Manual

1100 series standard, micro and preparative autosamplers
Table of Contents

Advertisement

Probable Causes
Suggested Actions
Troubleshooting and Test Functions

Initialization Failed

Initialization Failed
The autosampler failed to complete initialization correctly.
The autosampler initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined sequence. During
initialization, the processor monitors the position sensors and motor
encoders to check for correct movement. If one or more of the movements is
not successful, or is not detected, the error message is generated.
• Mechanical obstruction.
• Defective sampling unit flex board.
• Defective transport assembly flex board.
• Defective sampling unit motor.
• Defective ASM board.
❏ Ensure unobstructed movement of the transport assembly.
❏ Exchange the defective sampling unit motor.
❏ Exchange the transport assembly.
❏ Exchange the ASM board.