Error Information
Sampler Error Messages

Initialization Failed

Error ID: 34392, 34391
The sampler failed to complete initialization correctly.
The sampler initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined sequence. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Probable cause
1
2
3
4
5
174
Mechanical obstruction.
Defective needle station board.
Defective transport assembly flex board.
Defective hydraulic unit motor.
Defective main board.
Suggested actions
Ensure unobstructed movement of the
transport assembly.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.