Figure 15-6 Operation Event Register - Keithley 2182 User Manual

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

15-10
Additional SCPI Commands
Operation Event Register:
Bit B0, Calibrating (Cal) – Set bit indicates that the instrument is calibrating.
Bits B1 through B3 – Not used.
Bit B4, Measuring (Meas) – Set bit indicates that the instrument is performing a
measurement.
Bit B5, Trigger Layer (Trig) – Set bit indicates that the instrument is waiting in the Trigger
Layer of the Trigger Model.
Bits B6 and B7 – Not used.
Bit B8, Filter Settled (Filt) – Set bit indicates that the filter has settled.
Bit B9 – Not used.
Bit B10, Idle – Set bit indicates that the instrument is in the idle state.
Bits B11 through B15 – Not used.
Figure 15-6
Operation event register
Position
Event
Decimal
Weighting
Value
Value : 1 = Operation Event Set
Bit
B15 - B11
B10
B9
Idle
1024
10
(2 )
0/1
0 = Operation Event Cleared
B5
B3 - B1
B8
B7,B6
B4
Filt
Trig
Meas
256
32
16
8
5
4
(2 )
(2 )
(2 )
0/1
0/1
0/1
Events : Idle = Idle state of the 2182
Filt = Filter Settled
Trig = Trigger Layer
Meas = Measuring
Cal = Calibrating
B0
Cal
1
0
(2 )
0/1