Pulse Measurements Mode – Option 330
Pulse Measurement Settings

Pulse Top

Pulse Top is NOT the same as the pulse Peak. Pulse Top is used to set and
report the reference power level. You enter a value in percent, and the Pulse
Top value is reported in the Auto Analysis table in dB down from 100% of the
reference power level.
How to set Pulse Top
— Press Mrk ->/Tools.
— Then
— Then enter a percent value between 0 and 100 using the numeric keypad,
the

Grid

Enables and disables the display of a 10 x 10 grid on the FieldFox screen.
How to enable Grid display
— Press Mrk ->/Tools.
— Then Grid.

Trace Memory

A memory trace is a pulse measurement Tr 1 data trace (NOT the trace in the
Zoom window) that has been stored in the FieldFox volatile memory. It then
becomes static, meaning it is not updated with normal pulse measurements. It
can be displayed on the screen by itself, or with a pulse trace for comparison
purposes.
The Tr 1 data trace is displayed in yellow.
The memory trace is displayed in blue.
How to display data and memory traces
— Press Trace 6.
13-14
Pulse Top
.
arrows, or the rotary knob.
— When set to 10%, 10 dB down from the 100% reference level is
reported.
— When set to 50%, 3 dB down from the 100% reference level is
reported.
— When set to 90%, 0.5 dB down from the 100% reference level is
reported.
OFF
Grid OFF.
ON
A grid of X/Y lines appears on the screen.
Keysight N9912-90001 User's Guide