NA (Network Analyzer) Mode
NA Mode Settings
— Press Enter.
Single or Continuous Measure
This setting determines whether the FieldFox sweeps continuously or only once
each time the
or to allow you to save or analyze a specific measurement trace.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
— Press Sweep 3.
— Then choose one of the following:
— You can also use Run / Hold +/- to toggle between Single and Continuous.
Resolution (Number of Data Points)
Data points are individual measurements that are made and plotted across the
X-axis to form a trace. Select more data points to increase measurement
resolution. However, more data points also takes more time to complete an
entire measurement sweep.
When the Resolution is changed after a calibration is performed, the cal
becomes interpolated. Learn more in
How to set Resolution
— Press Sweep 3.
— Then Resolution.
— Then choose from the following:
101 |201
— Using SCPI, Resolution can be set to ANY number of points between 3 and
10001. See the Programming Guide at
http://na.support.keysight.com/fieldfox/help/Programming/webhelp/FFP
rogrammingHelp.htm
Keysight N9912-90001 User's Guide
Single
button is pressed. Use Single to conserve battery power
—
Single
Automatically sets Continuous OFF and causes FieldFox to
make ONE measurement, then hold for the next Single key press.
When a data trace is displayed, the entire trace is measured, then
holds. The Hold annotation changes to an arrow --> while the
measurement occurs.
—
Continuous
Makes continuous sweeps. This is the typical setting
when battery power is not critical.
|
401
|
601
|
801
"Interpolation" on page
|
1001
|
1601
|
4001
| 10001.