Battery Saver; File Folders - Keysight N9912A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9912A:
Table of Contents

Advertisement

System Settings
Preferences
How to setup Startup Mode
— Press System 7.
— Then Preferences.
— Then Preferences.
— Then scroll to Global Settings > Startup Mode and press
— Then repeatedly press the
choices.

Battery Saver

When the battery saver is OFF, the RF source power is left ON at the end of the
sweep which consumes more battery power. This feature is used in CAT, NA,
VVM modes.
How to set Battery Saver
— Press System 7.
— Then Preferences.
— Then

File folders

When enabled, FieldFox uses the last selected Device and Folder location as the
default device location after instrument power cycle.
Before setting the preferences for the File Folders, you must select the
Device and Folder locations. Refer to
page
18- 10
<Default>
is the factory default mode for your FieldFox model.
— Built-in Power Meter Mode is NOT allowed as a Startup Mode.
Battery
Saver.
— ON – Default setting. To conserve battery power, the RF source
power is turned OFF at the end of each CAT, NA, and VVM mode
sweep. This behavior does not impact FieldFox measurements.
— OFF – Source power is kept ON at all times, including during trigger
Hold mode and between single sweep acquisitions. The RF source
stays at the stop frequency until another sweep begins. This provides
a constant, uninterrupted RF output from the RF OUT port. This
behavior also provides the highest trace stability sweep-to-sweep
which may only be evident while using either single sweep or
Run/Hold. With Battery Saver OFF, the use of a full battery charge is
reduced by approximately 45 minutes.
17-1.
Startup Mode
to scroll through the available
Chapter 17, "File Management", on
Keysight N9912-90001 User's Guide
Ed
it.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents