Power Meter Mode (Option 302)
Power Meter Settings
You can also use Run / Hold +/- to toggle between Single and Continuous.
Step Detection Mode
The FieldFox supports the Step Detection feature that is present in U2000
series power sensors. When enabled, this feature reduces settling time after a
significant step in the measured power. For more information, please refer to
the USB Power Meter documentation.
How to set Step Detection mode
— Press Meas Setup 4.
— Then
— Then choose from the following:
Limits
In Power Meter Mode, Minimum and Maximum limits set a visible range of
acceptable readings between these two values. Readings that are below the
minimum or above the maximum are clearly visible on the power meter display.
How to set Limits
— Press Limit 8.
— Then choose from the following:
11- 8
Step Detection OFF
— ON Enables step detection (default setting)
— OFF Disables step detection
—
Min Limit OFF ON
Toggles the display of the Minimum limit ON and
OFF.
—
Min Limit Value
Sets the Minimum limit power value. Power reading
lower than this will exceed the limit.
—
Max Limit OFF ON
OFF.
—
Max Limit Value
Sets the Maximum limit power value. Power reading
higher than this will exceed the limit.
ON.
Toggles the display of the Maximum limit ON and
Keysight N9912-90001 User's Guide