15
Advanced Software Features

Ignore-Out-Of-Lock Mode

The Ignore Out Of Lock test mode enables all of the troubleshooting mode
functions, plus it causes the software to not check for an out-of-lock condition
before or during a measurement. This allows you to measure sources with high
close-in noise that normally would cause an out-of-lock condition and stop the
measurement. When Ignore Out Of Lock is selected, the user is responsible for
monitoring phase lock. This can be accomplished using an oscilloscope
connected to the test set Aux. Monitor port to verify the absence of a beatnote
and monitor the dc output level.
• When Ignore Out Of Lock is selected, the test system does not verify the
346
phase lock of the measurement. The user must ensure that the
measurement maintains phase lock during the measurement.