10
FM Discriminator Measurement Examples
7
.
e5505a_user_enter_source_tab
25 Jun 04 rev 2
Figure 202 Enter frequencies in Source tab
8
9
10
270
Choose the Sources tab from the Define Measurement window.
Enter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz).
a
Enter the same frequency for the detector input frequency. See
Figure
202.
Choose the Cal tab from the Define Measurement window.
Select Derive constant from FM rate and deviation as the calibration
method.
Set the FM Rate to 20 kHz and FM Deviation to 10 kHz, which are the
recommended FM rate and deviation. See
Figure 203
on page 271.
Agilent E5505A User's Guide