Approximate System Noise Floor vs. R Port Signal Level

The sensitivity of the phase noise measurement system can be improved by
increasing the signal power at the R input port (Signal Input) of the phase
detector in the test set.
the N5500A test set for a range of R input port signal levels from –15 dBm to
+15 dBm. These estimates of sensitivity assume the signal level at the L port is
appropriate for either the microwave or the RF mixer that is used (+7 dBm or
+15 dBm, respectively). The approximate phase detector calibration constant
that results from the input signal level at the R port is shown on the right side
of the graph.