Table 28. Parameter Data For The Microwave Source Measurement - Agilent Technologies e1420b User Manual

Phase noise measurement system
Hide thumbs Also See for e1420b:
Table of Contents

Advertisement

Table 28 Parameter data for the microwave source measurement
Step
1
2
3
4
5
Agilent E5505A User's Guide
Parameters
Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Minimum Number of Averages
FFT Quality
Sources Tab
Carrier Source
Frequency
Power
Carrier Source Output is
connected to:
Detector Input
Frequency
Reference Source
Frequency
Reference Source Power
VCO Tuning Parameters
Nominal Tune Constant
Tune Range ± Center Voltage
Input Resistance
Cal Tab
Phase Detector Constant
VCO Tune Constant
Phase Lock Loop Suppression
If Limit is exceeded
Block Diagram Tab
Carrier Source
Downconverter
Reference Source
Timebase
Phase Detector
Test Set Tune Voltage
Destination
VCO Tune Mode
Test Set Tab
Input Attenuation
LNA Low Pass Filter
LNA Gain
DC Block
PLL Integrator Attenuation
Absolute Measurement Examples
Data
Absolute Phase Noise (using a phase locked loop)
10 Hz
4 E + 6 Hz
4
Fast
12 E + 9 Hz
10 dBm
Test Set
600 E +6 Hz
600 E +6 Hz (same as Carrier Source Frequency)
16 dBm
40 E +3 Hz/V
± 10 Volts
0 Volts
600 Ω
Measure Phase Detector Constant
Calculate from expected VCO Tune Constant
Verify calculated phase locked loop suppression
Show Suppression Graph
Manual
Agilent N5502A/70422A
Agilent 8644B (System Control)
None
Automatic Detector Selection
Reference Source
DCFM
0 dB
20 MHz (Auto checked)
Auto Gain (Minimum Auto Gain –14 dB)
Not checked
0 dBm
6
199