To verify the PC digitizer card input's performance (continued) (continued)
Step
5 Select the FFT Analyzer Asset Control
e5505a_fft_control_panel
30 jun 04 rev 1
6 Configure the FFT Analyzer's Asset
7 Connect the PC digitizer adapter's
8 Click the Peak button.
9 Reconfigure the Function Generator.
10 Click the Peak button.
11 Close the FFT Analyzer's Asset
Agilent E5505A User's Guide
Panel.
Control Panel.
input connector to the output of the
function generator.
Control Panel.
PC Digitizer Performance Verification
Action
Path: System\Asset Control Panels\FFT Analyzer
Span: 0 to 2 MHz
Window Type: Flattop
Sweep: Continuous
Marker: Enable
See diagram in
step 1
on
The Frequency screen should reflect the 100 kHz
power specification: –29.00 dBv, ± 0.25 dBv
Frequency: 1.99 MHz
Amplitude: 100 mV peak to peak
The Frequency screen should reflect the 1.99 MHz
power specification: –29.00 dBv, ± 0.5 dBv
This concludes the procedure.