Instrument Removal

To remove an instrument from the system, use one of the following procedures.
Required tools
• #2 Phillips screwdriver
• #2 POZIDRIV screwdriver

Standard instrument

To remove an instrument from a rack

1 Turn off system power, but leave the system
2 Read "General Procedures and Techniques",
3 When all cables are disconnected (including
4 Slide the instrument out.
Agilent E5505A User's Guide
Step
computer turned on.
then disconnect all cables on the front and on
the rear panel.
the power cord), remove the screws in the
instrument's rack "ears" that hold it in the
rack.
Preventive Maintenance
Notes
If you do plan to turn computer power off
for any reason, have the computer system
administrator:
Shut down all running software.
Shut down the computer.
Most cables are fairly easy to remove and
reconnect, and have no special
considerations (besides making sure you
put the cables back in the right place).
Semirigid cables require more care,
especially when reconnecting them. Make
sure all semirigid cables, on the front and
back of an instrument are fully
disconnected before removing the unit.
If you feel any resistance when attempting
to pull the instrument out, STOP! Look
inside the cabinet and carefully examine
all surrounding cables. Make sure all
cables are fully disconnected.
21
461