Half-Rack-Width Instrument; To Remove A Half-Width Instrument From A System Rack - Agilent Technologies e1420b User Manual

Phase noise measurement system
Hide thumbs Also See for e1420b:
Table of Contents

Advertisement

21
Preventive Maintenance

Half-Rack-Width Instrument

To remove a half-width instrument from a system rack

1 Power off the system.
2 Remove the selected instrument's power cord
3 The instrument is attached to the half-rack width
4 Remove the power cord and other cables from the
5 Remove the four corner screws on the front of the
6 Slide both instruments, as a single unit, out from
7 Detach the lock links that secure the rear of the
8 Carefully and at the same time, push one
9 Store the "partner" instrument and lock links
462
from the power strip in the rack.
instrument beside it; remove that instrument's
power cord from the power strip also.
front and rear of both instruments.
rack panel that secures the instruments in place.
the front of the rack and set them on a secure, flat
surface.
instruments together by removing their screws.
instrument forward and pull the other back to
unhook the lock links that secure the front of the
instruments to each other.
while the selected instrument is out of the rack.
For details see the system installation
guide.
The instruments are secured together
by lock links at the front and rear. The
lock links at the rear attach with screws.
The lock links at the front hook together.
Note the location of cables for
re-installation.
The screws are located near the corners
of the face of the instrument.
Use a #2 Phillips screwdriver.
Use a #2 POZIDRIV screwdriver.
See
Figure 339
on page 463.
Only install the instruments as a pair;
individual installation is not secure.
Agilent E5505A User's Guide

Advertisement

Table of Contents
loading

Table of Contents